Maintenance of Electronic Compression Tester

Operators should familiarize themselves with the operating instructions, strictly observe the operating procedures, and do the following maintenance work:

1. Always keep the instrument clean. When it is not used for a long time, it should be covered with dust.

2. When the instrument fails, it should be checked and excluded by a professional. Do not run it sick.

3. After the instrument has been running for a period of time, an appropriate amount of lubricating oil should be injected into the filler hole on the front of the lifting sleeve. The surface of the lifting sleeve should be coated with an appropriate amount of grease.

4. Insist on periodic verification to maintain the excellent technical status of the instrument. The verification period is generally one year.

Electronic Compression Tester Troubleshooting and Maintenance

Symptoms Failure Analysis and Troubleshooting

Upper and lower pressure plate parallelism over tolerance 1. Loose sensor screws

2. Deformation of sensor elastomer

3. Loosen the upper platen screw to identify the fault location and perform corresponding repairs.

Excessive noise in the transmission system Reducer gear wear Replace gearbox

Limit protection device does not work Micro switch damage Replace micro switch